Siemens ULTRAMA6 Analyzers

ØHigh selectivity, can measure two Infrared channel components

ØWall mounting type at the scene(IP65)和rack mount type in the indoor(IP20)

ØRealize air tightness isolation between circuit section and analysis section in wall mounting type

ØWall mounting type was used for zones 1 and 2, according to ATEX 2G and ATEX 3G or FM/CSA Class I, Div. 2

ØHeating type analyzer, internal temperature of analysis section is 65℃

ØSIL2 Safety integrity LevelSafety integrity level certification

ØAnalog Output: 0/2/4mAActiveLoad Resistor≤750Ω

ØWarm-up Time: 30min; The maximal measurement accuracy can be achieved after 2h.

ØPreheating TemperatureExceed the room temperatureat least 35℃

ØSample Gas Flow: 18 ~ 60 l/h (0.3 ~ 1.0 l/min)

ØSample Gas Temperatuer: 0 ~ 50℃

ØSample Gashumidity: 90%RHno condensate

ØWorking Temperature: 5 ~ 45℃

ØWorking Humidty: 90%RH

ØWorking Pressure: 800 ~ 1200hPa (11 ~ 17psi)


标签: